Fault, Cause, Remedy - Dräger 8000 IC Instructions For Use Manual

Table of Contents

Advertisement

Fault, Cause, Remedy

Main module
Fault
Red Alarm LED is lit.
Alarm sound commences.
Red Inop. LED is lit.
Continuous tone commences.
Red Inop. LED is lit.
Continous tone commences,
also Err display in a module (except
air temperature control module).
N N N N
Red
LED is lit.
Continuous tone commences.
Cause
Fault in module.
Basic fault in electronics.
Fault in specific module:
Skin temperature control module/
humidity control module/O
control module
No mains supply.
Fault in incubator.
Alarm
Inop.
Check
Remedy
Fold down flap with Brief
Instructions for Use and identify
faulty module. See "Fault, Cause,
Remedy" of module.
Switch incubator off and on again.
If Inop. message not repeated:
press ƒ key and reset values.
If Inop. message repeated,
incubator not working.
Call DrägerService.
Switch off specific module.
All other functions will remain
2
operational.
Call DrägerService.
Check that mains plug is
connected to power supply.
Check that power is being supplied.
Inform in-house technical staff.
Call DrägerService.