Extending range of set values from 37 °C to 39 °C
Only do this if prescribed by a doctor.
The patient's core temperature must be monitored
very carefully.
1 Press unlocking key, 7 ,
2 yellow >37 °C control LED is lit.
The alarm limit for high temperature will be increased
from 38 °C to 40 °C.
3 Keep P or p key pressed until the value required
is displayed.
Set a value between 37 °C and 39 °C within 1minute.
Otherwise:
– >37 °C control LED will go out.
The lock against setting a higher range of values will
be in force again.
The alarm limit for high temperature will return to
38 °C again.
If a value below 37 °C is set after the range of values has
been extended:
– the extension is automatically cancelled.
If the actual temperature remains above 38 °C, the
high temperature alarm is triggered.
Recommendation to prevent an high temperature alarm:
Set value to 37.1 °C and wait until incubator has
cooled to 37.9 °C ,
then set value below 37 °C , as required.
Controlling air temperature
°C LUFT/AIR
2
°C
1
>37 °C
3
3
Check
Reset
Control
Operation
21