Checking that hand ports will stay closed
Open each hand port and then close carefully until
catch engages.
1 Grip rim of closed hand port and pull outwards – it
should not open.
If the hand port does not remain engaged:
Call DrägerService.
Checking that front door will stay closed
Open front door.
Lift up front door again.
2 Squeeze catches together on both sides, push front
door forward and release catches. Push firmly on
front door to ensure that catches fully engage. Red
mark on catches should no longer be visible.
If the front door does not engage:
Call DrägerService.
Checking air filter
Fold down back panel.
3 Open filter cover plate downwards.
If a filter is fitted:
Remove filter and check fitting date; label is on edge
of filter.
If filter is more than 2 months old:
Replace with a new filter.
Write fitting date on new filter label and stick to edge
of filter.
Press filter firmly into the seal.
Make sure that the direction of flow through filter is
correct. Arrow on filter has to point into incubator.
Close cover plate and lift back panel up again.
Testing tilting mechanism for bed
4 Raise left end of bed to maximum height.
5 Push down on raised bed with hand – bed must not
drop down.
If bed does drop down:
Call DrägerService.
Check tilting mechanism at right end of bed in the
same way.
Checking Readiness for Operation
1
1
2
3
5
4
Before each use
1
2
15