Dräger 8000 IC Instructions For Use Manual page 25

Table of Contents

Advertisement

When the temperature falls outside the 30 °C to 42 °C
measuring range:
– 3 dashes in the bottom of the display =
temperature below 30 °C
– 3 dashes in the top of the display =
temperature above 42 °C
see "Fault, Cause, Remedy" section, p. 43.
Activating skin temperature control
Allow at least 5 minutes for the skin temperature
sensor to reach the baby's temperature.
When value displayed remains constant,
1 Press ç key,
2 green Control LED is lit,
skin temperature control is in operation.
3 The display alternates between measured value of
skin temperature and the word SEt.
4 Flashing display of set value for skin temperature.
5 The measured value for air temperature continues to
be displayed.
The incubator offers an appropriate value, depending on
the situation.
Situation
Set value for air temperature not confirmed; 36.3 °C
Faulty or disconnected sensor
Actual skin temperature below 35 °C when 35 °C
switching over
Actual skin temperature between 35 °C
and 37 °C when switching over
Actual skin temperature above
37 °C when switching over
Confirming set value:
6 Press P or p key briefly, the display remains
continuously lit.
°C HAUT/SKIN/PEAU/PIEL
3
Set value
actual skin
temperature
selected
37 °C
°C HAUT/SKIN/PEAU/PIEL
6
Using skin temperature control
°C
°C
°C LUFT/AIR
2
5
°C
°C
4
Check
36 °C
±0,1
1
°C LUFT/AIR
°C
°C
6
Check
Control
36 °C
±0,1
Operation
>37 °C
Reset
Control
>37 °C
Reset
Control
25