Using humidity control

It is possible to control air temperature and relative
humidity in both automatic and manual modes.
In the automatic mode the set value for humidity is
calculated and adjusted automatically by the system,
depending on the air temperature set value, see
diagram. Maximum relative humidity is 75 %.
In the manual mode the set value for humidity can be
adjusted between 35 % and 85 %, using P or p keys.
Activating humidity control in automatic mode
1 Press ç key,
2 green Control LED is lit.
3 The word Aut (automatic) is displayed.
4 The word SEt and the measured relative humidity in
the incubator alternate in display.
5 Press P or p key briefly,
3 the word Aut remains continuously lit.
4 Measured relative humidity is displayed.
Activating humidity control in manual mode
6 Press P or p key again briefly to activate
manual mode.
7 Previous set value for humidity calculated by system is
displayed.
6 Enter set value required using P or p key.
At humidity values >60 % condensation may appear on
the incubator walls.
Then:
Decrease set value for humidity (if therapy allows).
Automatische Feuchteregelung
Automatic Humidity Regulation
100
90
80
70
60
50
40
30
20
10
0
28
29 30 31 32 33
34 35
Air Temperature set value
Lufttemperatur-Sollwert [°C]
% FEUCHTE/HUMID.
2
4
%
%
5
5
3
Control
1
% FEUCHTE/HUMID.
%
%
7
6
6
Control
Operation
Controlling humidity
36
37
38 39
29