Dräger 8000 IC Instructions For Use Manual page 39

Table of Contents

Advertisement

Canopy, inside and outside
Inner wall (folded down or removed)
Mattress
Bed and bed extension
Support brackets
Base plate
Trough
Incubator housing, inside and outside
Climate sensor:
Remove obvious soiling with disposable cloth and
detergent.
Wipe-disinfect surfaces.
Allow time for the disinfection process, then wipe the
surfaces again with a clean, damp cloth and dry.
Do not allow any liquid to get into climate sensor.
Fan
Seals
Hose seals
Connecting hose
Bottle holder or water container
Wash with detergent and rinse with clean water.
Disinfect in bath. Allow time for the disinfection
process, then rinse with clear water and dry;
or
sterilize at 120 °C (glove programme).
Use recommended cleaning and disinfecting agents only.
If other agents, such as alcohol for instance, are used
there is a risk of tension cracks in acrylic sheet and
Makrolon.
Do not spray or bath disinfect climate sensor.
Do not use UV-radiation on the incubator. This may cause
cracks in the acrylic sheet.
Care
Cleaning/disinfecting/sterilizing