Supplying Oxygen Manually - Dräger 8000 IC Instructions For Use Manual

Table of Contents

Advertisement

Operation

Supplying oxygen manually

Supplying oxygen manually
Switch off O
control module and wait for 3 minutes,
2
or Inop. alarm will occur.
Monitor O
concentration in incubator.
2
Use O
monitor, such as Dräger-Oxydig.
2
Preparation, p. 12.
To prepare flowmeter, see p. 12.
Supply oxygen from a medical gas pipeline system via
an O
flowmeter – push probe in fully.
2
When concentration required is reached, switch off
O
supply at flowmeter.
2
Recommended set values:
O
concentration
2
25 30 35 40
Vol.% (approx.)
O
flow
2
1.5 3.5 5.5 7.5
L/min
Using oxygen limiter
Preparation, p. 12.
The handwheel of the O
limiter has 2 settings.
2
Red setting:
no limitation of O
White setting:
O
supply limited to 6 L/min =
2
O
concentration about 40 vol.%, with
2
O
flow valve fully open.
2
If a lower O
set O
supply to less than 6 L/min.
2
Set handwheel as required.
34
45
50 55 60
9
11 13 15
supply.
2
concentration is required,
2
2