Care

Before re-using

Before re-using for a patient
Re-assemble equipment, see "Stripping down",
p. 37-38.
Check that the seals for support bracket in the trough
are movable.
Fitting seals in the hand port openings:
Open hand port.
Insert seal marked with an "L" on nose, into left
opening.
Insert seal marked with an "R" on nose, into right
opening.
Position nose of seal on hinge.
Put sealing lip (thin edge of profile) to the outside.
Close hand port.
Testing Readiness for Operation, see p. 14.
Allow water heater to run dry:
– to disinfect heater
– to test that "water shortage" display is working
properly.
Switch on humidity control.
Allow heater to run until "water shortage" alarm is
triggered, see p. 30.
Switch off humidity control.
Do not put full bottles into bottle holder until just
before placing baby in incubator.
If using water container:
Do not refill water container until just before placing
baby in incubator.
Operate incubator in standby, p. 19
or
switch off incubator, cover with dust cover and store
ready for use.