MaintenanceIntervals

Disposal of NiCd batteries

Maintenance Intervals

Clean and disinfect incubator and parts before
carrying out any maintenance procedures, and
before returning for repair.
O
sensor capsules
replace when sensors can no
2
longer be calibrated.
Disposal, see below.
Air filter
replace after two months, p. 15.
Dispose of as normal waste.
Seals of hand
replace, when the material
port openings
becomes brittle or tacky or
when a proper seal can no
longer be obtained.
Hose seals
replace, when the material
becomes brittle or tacky
and when the lamination is torn.
Ventilator motor
grease with 10 drops of
51524-HLP 32 oil every six
months by trained service
personnel.
NiCd battery for
replace once a year by
power failure
trained service personnel.
Disposal, see below.
Inspection and
every six months by trained
service*
personnel.
Disposal of O
sensor capsules and
2
NiCd batteries
– do not throw into fire; they may explode.
– do not open by force; risk of corrosion.
O
sensor capsules and batteries must be treated
2
as special waste:
in accordance with local waste disposal regulations.
Information can be obtained from local environmental and
public health authorities or from approved waste disposal
companies.
* According to DIN 31 051 the following definitions apply:
Inspection
= determining actual condition
Service
= measures to maintain required condition
Repair
= measures to re-establish required condition
Maintenance = inspection, service and, when necessary, repair