When there is a "skin temperature control" module
installed and it is in operation, the "air temperature control"
mode can be switched on again:
1 Press ç key,
2 green Control LED is lit.
The "air temperature control" mode is re-activated.
The set value for air temperature will be the last
value determined by the system.
3 The display alternates between the measured value for
air temperature and word SEt.
4 Confirm with P or p key and set a new value.

Using skin temperature control

When the skin temperature control option is installed.
Do not use on babies in shock as their skin temperature
is well below normal. Using skin temperature control
would increase the temperature of the incubator too
much.
To control air temperature, see p. 20.
Core temperature must be measured regularly.
Using skin temperature control
°C LUFT/AIR
2
3
°C
4
>37 °C
Reset
Control
1
Operation
Check
23