Operation

ThermoMonitoring

ThermoMonitoring, optional
For a better diagnosis of the child's thermal condition,
it is recommended that the temperature is measured both
centrally and peripherally. Both temperature measure-
ments may be displayed graphically via a Windows-PC or
a Dräger BabyScreen to document both the progress of
therapy for the child and the therapy itself.
Incubator must have:
– Second socket for peripheral skin temperature
sensor,
– Operator panel with © key,
– BabyLink interface and MediCable connecting cable
(optional).
ThermoMonitoring data may be displayed on:
– Babyguard 8000 patient monitor
– BabyScreen (optional)
– Windows-PC with ThermoView programme (optional).
Follow relevant Assembly and Operating Instructions.
Connecting peripheral skin temperature sensor
1 Connect white peripheral skin temperature sensor to
white socket on left of housing.
2 Route sensor cable through one of the hose seals.
3 Remove protective foil from adhesive pad and place
skin temperature sensor on pad.
Attach sensor to the patient's extremeties, preferably
on the foot, using adhesive pad.
Fix sensor cable in place with adhesive plaster.
Connect Windows-PC or BabyScreen to serial Baby-
Link interface using MediCable.
Follow relevant Instructions for Use.
Displaying measured values from peripheral
skin temperature sensor
4 Press © key and keep pressed. Yellow LED in key is lit.
5 Peripheral skin temperature is displayed.
4 Release © key,
5 skin temperature recorded centrally by first skin
temperature sensor is displayed again.
If 3 dashes appear in the display, see p. 25 and p. 27.
The measured value of the peripheral temperature does
not affect the control of the temperature in the incubator.
Both in air temperature and skin temperature mode the
peripheral skin temperature sensor can be disconnected.
No alarm will be given.
Both skin temperatures can be displayed when the
incubator is operating in the air temperature mode.
28
2
3
°C HAUT/SKIN/PEAU/PIEL
°C LUFT/AIR
°C
°C
5
4
°C
Check
Control
36 °C
±0,1
1
°C
>37 °C
Reset
Check
Control