If sensor plug is disconnected, climate sensor is
pulled out or sensor is faulty:
1 Red Alarm LED and
2 red sensor alarm LED are lit. Continuous tone
commences.
Then:
Connect climate sensor plug,
push back climate sensor or
change O
sensor capsules
2
without delay.
The intermittent tone can be suppressed for 10 minutes:
3 Press G key.
g
4 Yellow
LED and
1 red Alarm LED are lit.
For description of other alarms, see "Fault, Cause,
Remedy", p. 41 and 44.
Switching off O
control module
2
– when there is a fault in O
(Err display),
– when there is a sensor alarm which cannot be
rectified,
– when the deviation from set value remains above 5 %
even though O
sources have been switched off,
2
– when oxygen therapy is finished.
5 Press ç key.
6 Green Control LED is no longer lit.
Disconnect O
probe from terminal unit of medical
2
gas pipeline system or place in parking position.
If oxygen therapy is still required urgently:
Supply oxygen manually, see p. 34.
control module
2
Using oxygen control
Vol.%
O
2
Vol.%O
2
>40%
Cal.
2
Control
21%
VOL.%O
2
6
Vol.%
O
2
Vol.%O
2
>40%
Cal.
Control
21%
5
Operating
1
4
3
Check
33