Technical Data

Technical Data
Ambient conditions:
During operation
Temperature
Atmospheric pressure
Rel. humidity
During storage
Temperature
Atmospheric pressure
Rel. humidity
Operating data
Electrical power source
Heat output at 230 V
Air
Water heater
Current consumption at 230 V
Fuse for height adjustment
Performance data
Warming-up time
Temperature drop (at room temperature of 25 °C and
temperature inside incubator of 36 °C)
Two hand ports opened
Front door opened
Increase of O
oxygen concentration from 21 to 60 vol.%
2
Humidification
Air speed over the bed
Fresh air flow
CO
flushing, according to IEC 601-2-19/105.1
2
maximum CO
concentration inside incubator
2
Raising bed
Range of measured values
Air temperature
Skin temperature
O
concentration
2
Air humidity
Range of set values
Air temperature
Skin temperature
O
concentration
2
Air humidity
52
20 °C to 30 °C
900 to 1100 hPa
15 to 95 %
0 °C to 70 °C
900 to 1100 hPa
15 to 95 %
220 / 230 to 240 V AC (as per order)
50 / 60 Hz (as per order)
400 W
100 W
4.8 A
3.15 A IEC 127-2/III, 2 ea.
35 minutes from 20 °C to 31 °C
(with or without humidification)
<1 °C
<5 °C
<7 min
Evaporation of distilled or demineralised water
<8 cm/second
30 L/min
0.2 %
right or left, 10° angle of continuous tilt at both ends
10 °C to 50 °C
30 °C to 42 °C
0 to 99 vol.%
0 to 99 % r.H.
28 °C to 39 °C
35 °C to 37 °C
21 to 75 vol.%
35 to 85 % r.H.