Using oxygen control

Caution: physiological risks from oxygen
The air in the incubator should only be enriched with
oxygen if prescribed by a doctor. Oxygen enrichment
should be controlled on the basis of the arterially
measured oxygen partial pressure in the blood of
the patient. This is the only way in which both
hyperoxaemia (damage to the eyes) and hypoxaemia
(damage to the brain) can be prevented.
Connect probe on O
connecting hose to terminal
2
unit of medical gas pipeline system.
1 Press ç key.
2 Green Control LED is lit.
3 Yellow Cal. LED and
4 word CAL flash.
5 Set value 21 flashes in display.
Calibrate O
sensor within one minute, or sensor
2
alarm will occur.
Calibrating sensor
Calibrate sensor at outset of O
24 hours of continuous operation.
Pull climate sensor out of incubator at 90° angle.
Within one minute, or sensor alarm will occur:
6 Press 2 key.
7 Yellow Cal. LED is lit.
8 Display alternates between word CAL and "– –".
control and every
2