Dräger 8000 IC Instructions For Use Manual page 19

Table of Contents

Advertisement

Wait for warming-up time.
Wait for about 35 minutes before use.
Heat incubator in "air temperature control" operating
mode.
If incubator has to be used urgently, operate in standby
mode:
Set air temperature to 32 °C to 36 °C, see p.20.
Oxygen control off, see p.33.
Humidity control off, see p.30.
Just before placing the baby in the incubator:
Fill water bottles
If required use protective gloves.
1 Open back panel and fold down.
2 Disconnect empty water bottles and refill with 500 mL
clean, distilled or demineralised water (hospital
dispensary). Do not use any additives.
Refit filled water bottles and close back panel
again.
Then disinfect hands.
Use
– DIN 58363-IL 500-FL-BK1 or
– ISO 8536-1-IL 500-CL-HC1 Form A,
infusion bottles only.
If a water container is used:
Fill water container
If required use protective gloves.
1 Take cover of water container off and
2 fill container to mark with clean, distilled or
demineralised water (hospital dispensary).
Do not use any additives. Take care not to overfill.
Put cover of water container back.
Then disinfect hands.
Placing baby in the incubator
Open front door and pull out bed.
Put the baby on the bed and slide the bed back in
position.
Close the front door and check that the catches are
properly engaged.
Adjust the bed as required.
Adjusting working height
Adjust working height, as required, see p. 8.