Alarms
For deviations greater than ± 0.5 °C between set value
and actual value of skin temperature:
1 Red Alarm LED and
2 Red ±0.5 °C LED flash,
3 display flashes and intermittent tone commences.
The intermittent tone can be suppressed for 10 minutes:
4 Press G key.
g
5 Yellow
LED and
1 red Alarm LED are lit.
When measured value is within ±0.5 °C of the set value
again,
2 ±0.5 °C LED and
1 red Alarm LED go out.
intermittent tone ceases.
g
5 Yellow
LED goes out.
If the sensor plug is disconnected or sensor is faulty:
6 Display of 3 dashes in centre of display.
After 15 seconds:
Intermittent tone commences.
6 Display of 3 flashing dashes in centre of display.
7 Red Alarm LED and
8 red Sensor alarm LED are lit.
Then:
Connect sensor without delay or replace skin
temperature sensor.
The intermittent tone can be suppressed for 10 minutes:
Press G key, yellow
g
lit.
LED and red Alarm LED are
°C HAUT/SKIN/PEAU/PIEL
°C LUFT/AIR
°C
°C
3
°C
2
Check
Control
36 °C
±0,1
°C HAUT/SKIN/PEAU/PIEL
°C LUFT/AIR
°C
°C
6
°C
Check
Control
36 °C
8
±0,1
Operation
Using skin temperature control
Alarms
1
5
°C
4
>37 °C
Reset
Check
Control
1 7
°C
>37 °C
Check
Reset
Control
27