Performing Calibration - Canon CXDI-1 Operation Manual

Digital radiography system
Table of Contents

Advertisement

8.2 Performing Calibration

1
2
Start the calibration.
After setting the parameters on the calibration screen, touch [START].
The unit is ready to perform an x-ray exposure.
Preparing for Exposure Screen
Press the exposure button.
Press the exposure button on the x-ray generator. Data processing starts.
If data processing succeeds and two or more exposures are set, it is possible to proceed
with the next exposure.
Be sure to perform calibration exposures only for the number of set exposures.
Data Processing Success
– 99 –
8. Calibration
Continued on the following page >>

Advertisement

Table of Contents
loading

Table of Contents