Canon CXDI-1 Operation Manual page 109

Digital radiography system
Table of Contents

Advertisement

4
After the calibration, result will appear.
When calibration is successfully completed, the message "Calibration has
succeeded." is displayed. Touch [CHECK] to return to the initial screen.
When calibration has failed
A window is opened, and an error message is displayed.
After checking the error details, touch [OK] to close the window.
Touch [EXIT] to cancel calibration, or [RETAKE] to perform calibration again.
5
Change the exposure parameters.
If the settings of the x-ray generator were changed when starting calibration, change
exposure parameters other than number of exposures and bucky time.
The changed parameters will be saved when you touch [EXIT] or change the sensor unit.
6
Return to the menu screen.
Touch [EXIT].
7
Return the x-ray generator to the settings for normal exposure.
Be sure to change the settings of the x-ray generator to those for normal exposures.
REX check
If exposure is performed with the settings for calibration exposure, the REX will
display a value differing significantly from normal exposure, and a normal image will
not be obtained.
It is recommended that you always check the REX.
Calibration Successful
– 101 –
8. Calibration