9. Self test

9.1 Self test using a sensor unit with a removable grid unit

This procedure checks the image data transfer path of the CXDI system and individual sensor
performance within the sensor unit.
1
This displays the Self Test Screen.
From the exposure screen, touch [SYSTEM] > [SELF TEST].
2
Select the sensor unit.
If multiple sensor units can be used, first select a sensor unit by touching that sensor unit
button.
3
Remove the grid unit.
If a sensor unit with removable grid unit is selected in step 2, remove the grid unit.
If [TEST START] is touched while the grid unit is attached, it will display an error, and
the self test cannot be performed.