Edge Enhance - Canon CXDI-1 Operation Manual

Digital radiography system
Table of Contents

Advertisement

7. Processing the Image (QA Mode Screen)

7.5 EDGE ENHANCE.

NOTE:
When parameters are set to excessively high values, artifacts may be
generated or effective image information may be degraded.
If you think the image looks as if it has been over-processed, set the
parameters to lower values, or void the process.
The edge enhancement process emphasizes the edges of the image and sharpens the image.
However, it is difficult to check the effect of changes by modifying the parameters in the QA
screen.
1
Enable edge enhancement.
Touch and highlight (white character) [EDGE ENHANCEMENT].
2
Set the degree of enhancement.
Touch
or
to select a setting from 1 to 11.
The displayed value on the left is the saved value, and the value on the right is the
modified value.
The effect is proportional to the setting value.
Higher values result in higher emphasis, and lower values result in lower emphasis.
3
Set the enhancement frequency.
Touch
or
to select a setting from 1 to 11.
The displayed value on the left is the saved value, and the value on the right is the
modified value.
The larger the value, the clearer the definition of detailed (high frequency) structures, such as
bone trabeculae. Low frequency values result in clearer definition of larger (low
frequency) structures, such as soft tissue.
The settings and corresponding frequencies are shown below:
Table 7.5.1: Setting Values – Enhancement Frequency
Setting
FREQUENCY
(lp/mm)
EDGE ENHANCE Screen
1
2
3
4
0.06 0.08 0.11 0.15 0.20 0.27 0.36 0.47 0.69 0.81 1.28
– 76 –
5
6
7
8
9
10
11

Advertisement