11. Questions and Answers

11.2 Entering/Changing Data

There are exposure parameters that are not displayed on the exposure screen.
Q2-1
You can select the exposure parameters to be displayed from the SYSTEM SETUP screen (→ Setup
A2-1
Guide).
Parameters set to off are not displayed.
Do you have to enter all the data on a patient?
Q2-2
An error occurs if any of the required fields (→ Setup Guide) are not completed.
A2-2
The required fields can be selected.
How can patient data be changed after input?
Q2-3
Touch [PATIENT]. Now the data can be changed. The [OK] at the bottom right of the software keyboard
A2-3
changes to [CHANGE]. To exit, touch [CHANGE].
How can exposure parameters be changed on the category type screen?
Q2-4
Unlike the tray type screen, they cannot be changed directly from the category type screen. Touch
A2-4
[CHANGE] to display the screen for changing parameters.
Although the exposure parameters have been changed, they returned to the previous settings
Q2-5
after an exposure was performed.
Unless you touch [SAVE PARAM] and save the parameters after changing them (→ page 18), the changed
A2-5
parameters will return to the previous settings after an exposure has been taken.
There is a parameter that cannot be changed during the exposure operation.
Q2-6
For msec and mAs, only the one that has been selected on the SYSTEM SETUP screen can be changed.
A2-6
Selection can be made in the Exposure Condition Display (→ Setup Guide).
Exposure parameters cannot be changed freely.
Q2-7
If [ENABLE] has been selected for PARAMETER TABLE (→ Setup Guide) on the SYSTEM SETUP
A2-7
screen, only the preset values in the exposure parameter table will be displayed. Setting to [DISABLE]
allows setting of any user-selected value.