Page 2
This manual describes the cleaning, disinfection, and sterilization procedures for the ultrasound transducer. For the operating precautions and procedures for the transducers, refer to the operation manual for each transducer. Trademarks This manual may include trademarks or registered trademarks of companies other than Canon Medical Systems. 2B702-068EN*N...
1. Safety Precautions Meaning of Signal Words DANGER WARNING CAUTION In this manual, the signal words , and are used regarding safety and other important instructions. The signal words and their meanings are defined as follows. Please understand their meanings clearly before reading this manual. Signal word Meaning Indicates an imminently hazardous situation which, if not...
Page 4
4) After gas sterilization, degas the transducer. Gas residues on the transducer may be harmful to the human body. 5) The efficacy of the cleaning solutions, disinfectants, and sterilizing gases is not guaranteed by Canon Medical Systems. Contact the manufacturers for information on the activity of the products.
Page 5
Chemical Hazard (for USA only) Observe the following instruction in order to protect patients and operators from inflammation or poisoning by chemical substances. This product can expose you to chemicals including lead and lead WARNING: compounds, which are known to the State of California to cause cancer, and phthalates, which are known to the State of California to cause birth defects or other reproductive harm.
2. Cleaning, Disinfection, Sterilization This section describes the methods for cleaning, disinfection, and sterilization. Cleaning and disinfection between patients must be performed to prevent transmission of disease. All transducers must be thoroughly cleaned prior to disinfection. The level of disinfection required is based on patient contact. Cleaning <<Items to be used: Protective gloves, cleaning solution or cleaning wipes, purified water, clean soft cloth or gauze, single-use sponge*>>...
Page 7
Disinfection Before disinfection, the transducer must be cleaned. Note that local regulations may require that the transducer be disinfected before sterilization. <<Items to be used: Sterile protective gloves, disinfectant, sterile water or deionized water, sterile soft cloth or gauze>> (1) Wear sterile protective gloves to prevent infection. Wear new sterile protective gloves each time disinfection is performed.
Page 8
NOTE 1. When multiple model names included in the List of Chemicals are referred to, note that they are abbreviated. For example, "PSU-25BT and PSU-30BT" is abbreviated as "PSU-25BT/30BT". 2. For the transducers supported by each diagnostic ultrasound system, refer to the operation manual supplied with the diagnostic ultrasound system.
Page 9
5 min. 10 min. Temperature Room temperature Room temperature Humidity Normal humidity Normal humidity Model name Pressure Normal pressure Normal pressure Immersible range PSU-25BT/30BT/70BT PVU-375BT/382BT Type A PSU-50BT Type A PVU-712BT PVU-674MV Type A – – – PLU-704BT Type A...
Page 11
Room temperature Room temperature Room temperature Humidity Normal humidity Normal humidity Normal humidity Model name Pressure Normal pressure Normal pressure Normal pressure Immersible range PSU-25BT/30BT/70BT PVU-375BT/382BT Type A PSU-50BT Type A PVU-712BT PVU-674MV Type A – – – PLU-704BT Type A...
Page 13
Humidity Normal humidity Normal humidity humidity humidity Normal Normal Model name Pressure Normal pressure Normal pressure pressure pressure Immersible range PSU-25BT/30BT/70BT PVU-375BT/382BT Type A PSU-50BT Type A PVU-712BT PVU-674MV Type A PLU-704BT Type A PVU-745BTV Type A PVU-745BTF/745BTH/770ST Type A...
Page 15
5 min. Wiping only Temperature Room temperature Room temperature Humidity Normal humidity Normal humidity Model name Pressure Normal pressure Normal pressure Immersible range PSU-25BT/30BT/70BT PVU-375BT/382BT Type A PSU-50BT Type A PVU-712BT PVU-674MV Type A – – – PLU-704BT Type A...
Page 17
90 min. CAN: Canada Temperature Room temperature 25°C (77°F) Humidity Normal humidity Normal humidity Model name Pressure Normal pressure Normal pressure Immersible range PSU-25BT/30BT/70BT PVU-375BT/382BT Type A PSU-50BT Type A PVU-712BT PVU-674MV Type A – PLU-704BT Type A PVU-745BTV Type A...
Page 18
Ortho- Glutaraldehyde Hydrogen peroxide phthalaldehyde Revital-Ox™ ® ® Trophon Resert ® ® ® ® ® WAVICIDE Sporicidin CIDEX Trophon SPOROX Ⅱ High Level Disinfectant Medical Chemical Contec Inc. J&J Nanosonics Limited Sultan Healthcare STERIS Corporation Corporation 1.12% Use the disinfectant cartridges 2.5% 0.55% 7.5%...
Page 19
(temperature, humidity, time) to be set by the Humidity pressure, or time) to be operator. set by the operator. Model name Pressure 980 hPa [gauge] PSU-25BT/30BT/70BT *7, *8 – – PVU-375BT/382BT PSU-50BT – – – – PVU-712BT PVU-674MV –...
4. Other Information ● For Europe CANON MEDICAL SYSTEMS EUROPE B.V. Zilverstraat 1, 2718 RP Zoetermeer, The Netherlands https://eu.medical.canon For other countries, please contact your local distributor. ● Legal manufacturer Canon Medical Systems Corporation 1385, Shimoishigami, Otawara-shi, Tochigi 324-8550, Japan...
Page 31
室 温 室 温 室 温 室 温 せん。 湿 度 常 湿 常 湿 常 湿 常 湿 圧 ⼒ 常 圧 常 圧 常 圧 常 圧 機種名 浸漬可能範囲 PSU-25BT/30BT/70BT PVU-375BT/382BT/770ST Type A PLU-704BT PVU-674BT Type A PLU-1005BT/1204BT PVL-715RS Type A – – PSU-50BT Type A PVU-712BT PVU-674MV Type A – PVU-745BTV Type A –...
Page 39
PVU-674BT, PLU-1005BT A 类 PVU-781VT/781VTE A 类 PLU-1204BT A 类 PVU-681MVL A 类 PVU-350BTP A 类 PLU-308BTP A 类 PLU-1202BT A 类 OK : 允许使用化学品 – : 不允许使用化学品 *1: 注意,当地法规可能要求换能器在灭菌前进行消毒。 *2: 只有当换能器不被内部使用。 ● A 类 (PSU-25BT etc.) 标签 浸没范围 切勿将换能器的非防水部分浸没在水或清洗溶剂等液体中。浸没可能导致触电。 2B702-068EN*N...