Error
Error Message (Upper box: Error type
Code
Capture Board Error
-132
Capture Board Error. Call a service engineer.
Capture Board: Sequence Error
-134
Failed to synchronize with the sensor unit. Call a service engineer.
Capture Board: Sequence Error
-135
X-ray monitor in sensor unit failed to detect the beginning of irradiation.
Capture Board: Data Transmit Error
-136
Failed to transmit the data. Call a service engineer.
Capture Board: Control Line Error
-137
Control line error has occurred in capture board. Call a service engineer.
Sensor Unit: Sensor Analogue Power Supply Error
-138
Sensor Analogue Power Supply was detected. Call a service engineer.
Grid ID Readout Error
-139
Grid ID readout error has been detected.
Acquisition Parameter Error
-140
Detected Illegal Parameter Error.
LUT Write Error
-141
Quit the exam and perform calibration.
Calibration Table Error
-143
Perform Calibration.
Sensor Unit Configuration Error
-144
Incorrect Sensor Unit Configuration. Call a service engineer.
Sensor Unit: Firmware Write Error
-145
The firmware is incorrect.
Capture Device Access Error
-146
Detected Communication Error
Either an error occurred in the sensor, or the sensor power may be turned off.
Capture Device Access Error
-147
Detected Communication Error
Either an error occurred in the sensor, or the sensor power may be turned off.
Sensor Unregistered Error.
-148
The connected sensor cannot be used because it is unregistered.
Lower box: Error description)
– 115 –
10. Before Calling a Service Engineer
Remedy
Touch [OK] to close the window, and then see
1.2 Shutting Down the CXDI (→ page 5),
and turn off the CXDI.
When these errors occur, note the error code
and contact a Canon representative or
distributor.
Touch [OK] to close the window. Exposure can
be continued.
When these errors occur, note the error code
and contact a Canon representative or
distributor.
Touch [OK] to close the window, and then see
1.2 Shutting Down the CXDI (→ page 5),
and turn off the CXDI.
When these errors occur, note the error code
and contact a Canon representative or
distributor.
Touch [OK] to close the window, and then see
8. Calibration (→ page 97) to perform
calibration.
Touch [OK] to close the window, and then see
1.2 Shutting Down the CXDI (→ page 5),
and turn off the CXDI.
When these errors occur, note the error code
and contact a Canon representative or
distributor.
If sensor power is turned off, turn the sensor
power on and try performing the operation
again.
If sensor power is on, note the error code and
contact a Canon representative or distributor.
Touch [OK] to close the window, and then
connect a registered sensor.
Continued on the following page >>