Settings - Canon CXDI-1 Operation Manual

Digital radiography system
Table of Contents

Advertisement

11.3 Settings

A new exposure mode button cannot be made.
Q3-1
To create a new exposure method, copy the exposure method used before and then edit it. (→ Setup
A3-1
Guide)
This eliminates the need to enter the parameters from the beginning.
What is a bucky msec?
Q3-2
It is a measure of the time for linking the movable grid with the X-ray irradiation. When it is not set
A3-2
properly, vertical lines may appear on the image.
What is the difference between the fixed ROI and phototimer?
Q3-3
In the Fixed ROI settings, the adjustment range for LUT processing (→ page 70) is set. By setting the
A3-3
phototimer, the position of the phototimer installed in the CXDI is set.
What is an X-ray monitor?
Q3-4
It is used to detect the completion of X-ray irradiation and judge the timing for image generation.
A3-4
Normally, this cannot be changed. (→ Setup Guide)