10.4 Warning

Error
Code
SMART Error
-9
An error is detected in HDD.
It's possible that the HDD(1 or 2) will break down. Call a service person.
HDD Communication Error
Cannot communicate with HDD.
-10
Mirroring function will be disabled since the HDD(1 or 2) is not available. Call a
service person.
SMART Error
An error is detected in HDD.
-11
It's possible that the HDD(1 or 2) will break down and the data will be lost. Call
a service person.
Sensor Unit: SelfCheck Error
-150
Detected an Error.
Sensor Unit: Flash ROM Error
-151
Detected Flash ROM Error. Call a service engineer.
Sensor Unit: Sequence Error
-152
This image may have a problem.
Sensor Unit: Temperature Warning
-153
The internal temperature of the Sensor Unit is too high!
Do not leave the Sensor Unit in READY condition.
Capture Board Error
-160
Detected Capture Board Error.
Capture Board Error
-161
Capture Board Error.
Capture Board: Sequence Error
-162
X-ray monitor in sensor unit failed to detect the beginning of irradiation.
Capture Board: Sequence Error
-163
X-ray monitor in sensor unit failed to detect the end of irradiation.
Image quality attention
-164
CXDI has stopped irradiation. Check image.
DICOM Connect Error
Cannot connect to target. Check network or port number setting.
-502
(The following message is added when transferring all data at once.)
If operating offline, press the Close button and check the output status of the
Process Viewer.
DICOM Transfer Error
-503
Error has occurred during association.
Error Message (Upper box: Error type
Lower box: Error description)
10. Before Calling a Service Engineer
– 119 –
Remedy
Touch [OK] to close the window, and then see
1.2 Shutting Down the CXDI (→ page 5),
and turn off the CXDI.
When these errors occur, note the error code
and contact a Canon representative or
distributor.
Touch [OK] to close the window. As image
quality cannot be guaranteed, take an image
again.
Touch [OK] to close the dialogue window.
Lower the temperature of the room where the
sensor unit is installed, or wait until the
temperature in the sensor unit drops.
Touch [OK] to close the window, and then see
1.2 Shutting Down the CXDI (→ page 5),
and turn off the CXDI.
When these errors occur, note the error code
and contact a Canon representative or
distributor.
Touch [OK] to close the window. As image
quality cannot be guaranteed, take an image
again.
• After confirming that the network is normal,
touch [RETRY].
• Touch [CANCEL] to close the window, and
then see Setup Guide and set the
parameters.
• If running the software offline, check that
the output status of the process viewer is
[IDLE]. If output status is [SEND], change it
to [IDLE].
Continued on the following page >>