8. Trouble Shooting

8.1

RadEye doesn't power on

8.1.1

Press "On" for at least one second

Keep the button " " pressed until the LED and the sound is
active. If the button is pressed only shortly, then the start up
routine is not activated. The LED may flash then only very
shortly.
8.1.2

Check battery

If the RadEye does not respond to a long press (> 1s) of the
" " button, then please use the following step by step proce-
dure:
A) Remove the current batteries
B) Clean the battery contacts
C) Check the required orientation of the batteries according to
the printed "+" symbols in the battery compartment.
D) Insert two -"AAA" with sufficient capacity: Rechargeable
AAA batteries should be completely loaded, non rechargeable
batteries should supply together at least 2.3 Volt under load.
E) Press the " " button for more than one second (see also
8.1.1)
RadEye G20
Iw/Tr/Ff 26.11.2012
DB-067 E
8-1