Trouble Shooting; Radeye Sx Doesn't Power On; Press "On" For At Least One Second; Check Battery - Thermo Scientific RadEye GX Operating Instructions Manual

Multi-purpose survey meter for external gm-detectors
Table of Contents

Advertisement

10. Trouble Shooting

10.1

RadEye SX doesn't power on

10.1.1

Press "On" for at least one second

Keep the button "" pressed until the LED and the sound is ac-
tive. If the button is pressed only shortly, then the start up routine
is not activated. The LED may flash then only very shortly.
10.1.2

Check battery

If the RadEye SX does not respond to a long press (> 1s) of the
"" button, then please use the following step by step procedure:
A) Remove the current batteries
B) Clean the battery contacts
C) Check the required orientation of the batteries according to the
printed "+" symbols in the battery compartment.
D) Insert two -"AAA" with sufficient capacity: Rechargeable
AAA batteries should be completely loaded, non rechargeable
batteries should supply together at least 2.3 Volt under load.
E) Press the "" button for more than one second (see also 2.4)
RadEye GX, GX-L
Pi/Ff 14.09.2016
DB-074 E
10-1