Trouble Shooting; Radeye Doesn't Power On; Press "On" For At Least One Second; Check Battery - Thermo Scientific RadEye B20 Operating Instructions Manual

Multi-purpose survey meter
Hide thumbs Also See for RadEye B20:
Table of Contents

Advertisement

8. Trouble Shooting

8.1

RadEye doesn't power on

8.1.1

Press "On" for at least one second

Keep the button "" pressed until the LED and the sound is
active. If the button is pressed only shortly, then the start up
routine is not activated. The LED may flash then only very
shortly.
8.1.2

Check battery

If the RadEye does not respond to a long press (> 1s) of the
"" button, then please use the following step by step proce-
dure:
A) Remove the current batteries
B) Clean the battery contacts
C) Check the required orientation of the batteries according to
the printed "+" symbols in the battery compartment.
D) Insert two -"AAA" with sufficient capacity: Rechargeable
AAA batteries should be completely loaded, non rechargeable
batteries should supply together at least 2.3 Volt under load.
E) Press the "" button for more than one second (see also
8.1.1)
RadEye B20/-ER
Iw/Tr/Ff 06.03.2017
DB-065 E
8-1