5.4

Low Battery warning

If the battery voltage is below the configured threshold (see
chapter 4.15) the following warning appears:
The beeper generates a single pulse every 32s. This acoustic
warning can be suppressed by pressing the alarm acknowledge
key. After 8h this warning comes up again. The battery needs
to be changed. However, the RadEye can still be operated for
several hours.
If the battery voltage falls below 2V, the battery symbol starts
flashing. An acoustic warning is generated every 8s and can't
be acknowledged. The battery needs to be changed as soon as
possible.
5-4
DB-066 E
RadEye PRD-ER
Iw/Tr/Ff 06.07.2015