AccuPyc II 1340
Unit[n]
SN1234 10 cm3
Analyze
Chamber insert?
None
Unit[n]
SN1234 10 cm3
Analyze
[ENTER] to start
[Escape] to cancel
Calibrate
This command enables you to perform a full volume slope and offset calibration; enter a temperature
or pressure value obtained from an external reference device; copy to or load from a USB stick; and
reset the pressure calibration.
.
Press
+
(decimal) to access the Calibrate function.
ALT
Unit[n]
SN1234
10 cm3
X - X - X
Calibration Type?
Volume
Aug 2013
Enables you to choose the insert you are using.
Choices: None, 0.1 cm
None, 3.5 cm
3
None, 35 cm
None, 650 cm
pycnometer)
Refer to
MULTIVOLUME
additional information on MultiVolume inserts.
Press
CHOICE
until the appropriate insert is displayed or
None if an insert is not being used, then press
Choose one of the following:
• Press
to start the analysis.
ENTER
The analysis begins and operational status messages are
continually displayed during analysis.
• Press
to cancel the analysis.
Alt + CLEAR
• Press
to store the data you entered and return to dis-
SAVE
play mode.
Choices:
Volume, Temperature, Pressure,
Copy to USB Stick, Load from USB Stick,
Reset Pressure Calibration
Press
until the desired option is displayed:
CHOICE
Press
.
ENTER
3
3
(for 1 cm
unit)
3
3
3
, 1.0 cm
(for 10 cm
unit)
3
3
, 10.0 cm
(for 100 cm
3
3
, 1300 cm
(for 2000 cm
INSERTS, page
J-1
ENTER
System Commands
unit)
3
for
.
3-25