Appendix A
MAN_ERR:
Pressure overrange
Cause:
Action:
No A.C. Found
Cause:
Action:
No collected data to report,
or all cycles excluded
Cause:
Action:
Printer port
not responding
Cause:
Action:
Queuing print job
[ESCAPE] to cancel
Cause:
Action:
A-8
The fill valve was left open until the maximum system pressure was
exceeded.
Close the fill valve and open the vent and expansion valves. Allow the
pressure to stabilize.
The control module cannot detect an analysis module.
Contact your Micromeritics service representative.
You requested a report in which there is either no data available or data
have been excluded via review mode.
Initiate an automatic operation or return to review mode and include at
least one cycle.
Your printer cable is not connected or the printer is not turned on.
Check to make sure the printer is properly connected to the
pycnometer, is turned on, and is on line.
Status message displayed when a report is being printed.
None; this is a status message only.
AccuPyc II 1340
Oct 2013