AccuPyc II 1340
Reset to factory
defaults
Cause:
Action:
Sending line (line number)
[ESCAPE] to cancel
Cause:
Action:
Transmission port
not responding
Cause:
Action:
Transmission port
waiting for Xon
Cause:
Action:
Oct 2013
.
The
(period) key was pressed at startup and the instrument was reset.
Data files and HTML reports have been erased. Setup options have
been returned to default values.
Review setup options.
Status message displayed when data are being transmitted.
None; this is a status message only.
The receiving device took longer than five seconds to acknowledge
receipt of data from the pycnometer.
Make sure the receiving device is properly connected to the
pycnometer RS-232 port and is turned on. Verify that the serial I/O
parameters controlling the receiving device correspond with the data
transmission parameters in the set up mode.
The receiving device stopped transmission by sending an Xoff, and
hasn't resumed the transmission by sending an Xon.
None; when the receiving device is ready for more data, it should send
the pycnometer an Xon.
Appendix A
A-9