Calibrating Transducers and Servo Valves
Calibrating Transducers and Servo Valves
Calibrate the 1000-Torr Pressure Transducers
Verifying the 1000-Torr transducers' accuracy requires the use of:
• 004-09814-00 Paroscientific Pressure Calibration Standard OR
• 004-09603-00 Crystal Pressure Gauge (if 004-09814-00 is
• 003-09623-01 Vacuum Calibration Standard. (Yellow Jacket)
1. Ensure that nitrogen is attached to the Adsorptive gas inlet port on the instrument and the
pressure is set to 15 psig.
2. Pressurize the ports.
a. Select Unit [n] > Show Instrument Schematic, Unit [n] > Enable Manual Control.
b. Close all valves (double-click a valve to open or close it).
c. Open valves 4-b and 4-s. Open valve 2 (adsorptive) for 15 seconds.
d. Close all valves.
48
Part 4. Operational Verification
unavailable)
Gemini VII 2390 Confirm Installation Instructions and Checklist
Gemini VII 2390 Confirm
Feb 09