6. Press
until the average measured volume is displayed; record the value.
CHOICE
7. Press
to access Setup.
Alt + 2
8. Press
CHOICE
until Calibration Data is displayed.
Unit[n]
SN1234
Set-up Type?
Calibration Data
9. Press
ENTER
; the Chamber Insert prompt is displayed.
10. Ensure that None is selected, then press
11. Subtract the average volume of the 0.1-cm
volume displayed in the prompt; record this value.
12. Press
ENTER
to display the Expansion Volume prompt; record the value displayed.
13. Record the value displayed for the expansion volume, then press
to the Calibration Data prompt.
14. Press
; the Chamber Insert prompt is displayed.
ENTER
15. Press
CHOICE
until 0.1 cm
prompt. Enter the value you recorded in Step 11.
16. Press
to display the Expansion Volume prompt; enter the value you recorded in
ENTER
Step 12.
17. Press
to save the information you entered and return to the Reload prompt.
SAVE
The pycnometer is now ready for analyses using the 0.1-cm
10 cm3
3
is displayed, then press
to display the Cell Volume prompt.
ENTER
3
insert/cup (recorded in Step 6) from the cell
Enter
to display the Cell Volume
3
insert/cup.
. This returns you
ENTER