AccuPyc II 1340
USR_ERR:
No data to review
Cause:
Action:
USR_ERR: Number of
cycles must be = 5
Cause:
Action:
USR_ERR: Out of range
Cause:
Action:
USR_ERR: Pressure overrange
Cause:
Action:
Valve failure
[ENTER] to retry.
Cause:
Action:
Oct 2013
You tried to review data for an automatic operation when there were no
data to review.
Abandon request.
You tried to enable run precision without increasing the number of runs
to at least five.
Increase the number of runs to five or abandon request.
You tried to enter a value that is out of the valid range.
Enter a value in the specified range (refer to Chapter 4).
A pressure overrange occurred but was left uncorrected.
Return to manual mode and vent the system.
One (or more) of the system valves failed to operate when the
instrument was initialized.
Press
ENTER
to retry. If still unsucessful, contact your service
representative.