System Commands
Unit[n]
SN1234
10 cm3
Analysis Parameters
Percent Full-Scale:
0.05%
Report Options
The Report options command enables you to specify the report mode (density or volume for analysis)
and the report destination.
Unit[n]
SN1234
10 cm3
Set-up Type?
Report Options
Unit[n]
SN1234
10 cm3
Report Options
Anls Display Mode?
Density
Unit[n]
SN1234
10 cm3
Report Options
Request Sample ID?
Yes
3-14
This prompt displays when you selected Yes for run precision
and if you have requested more than five cycles.
Enter the run precision volume tolerance which is expressed
as a percent of nominal cell volume (sample capacity), then
press
ENTER
.
Range:
0.01% to 50%
Press
CHOICE
until Report Options is displayed, then press
ENTER
.
Press
to select Density or Volume, the mode in
CHOICE
which the analysis report will be displayed on the
pycnometer.
If you select Density, a Sample Mass? prompt displays
during the Analysis function so that you can enter a value for
the sample mass. For accurate data, the sample mass must be
entered when choosing Density.
Press
.
ENTER
A sample ID (identification) is a unique identifier of the
sample. Using sample IDs can help you keep track of data
from multiple analyses. For example, you can use the
sample ID as a date and time stamp. A Sample ID prompt
displays during the analysis operation, allowing you to enter
the identification when you select Yes to this prompt.
Choices: Yes, No
Press
to select Yes or No, then press
CHOICE
AccuPyc II 1340
.
ENTER
Aug 2013