Micromeritics AccuPyc II 1340 Operator's Manual page 103

Table of Contents

Advertisement

AccuPyc II 1340
ANLSERR: Underrange
NN/ZZ cycles completed
Cause:
Action:
Automatic operation
has been canceled
Cause:
Action:
CAL_ERR: Reset Pressure Cal. ?
Cause:
Action:
Action:
CAL_ERR: Temperature
Calibration Failed
Cause:
Action:
Oct 2013
A pressure underrange occurred during an analysis because an error
occurred in the pressure measurement electronics.
Call a Micromeritics Service Representative.
The automatic operation has been canceled by the user.
Wait for the termination process to complete or end the termination
process by pressing
Alt + CLEAR
The calibration information could not be read from the command or
analysis module.
Yes
Enter
Choose
and press
calibration to nominal values. Resetting to nominal may restore
normal readings but a proper calibration should be performed by a
service technician. See the section Reset Pressure Calibration in the
System Commands section.
No
Choose
to avoid resetting the calibration. The instrument will be
unresponsive.
The temperature calibration was unsuccessful.
Contact your service representative.
two more times.
to reset the pressure and temperature
Appendix A
A-3