4.8.2 Switching on the maintenance condition

Switch on the maintenance condition manually when you want to clean, service
or repair (remove and replace) an IQ sensor.
1
Call up the measured value display with
2
Open the Display/Options menu with
<
3
Using

condition on/off, and confirm with
A window that informs you about the maintenance condition opens up.
<OK>
4
Use
to confirm Continue. The selected sensor is in the
maintenance condition. Linked outputs are frozen.