Recorder - Xylem WTW DIQ/S 281 Operating Manual

Universal transmitter for one digital sensor
Hide thumbs Also See for WTW DIQ/S 281:
Table of Contents

Advertisement

Outputs
Function
Settings
106
<
9
Using

The new settings are stored.

6.7.1 Recorder

The measured values of the linked sensor at the current output are set up as cur-
rent intensity in the Recorder application. The output of measured values is laid
down in the Recorder type, Start value and End value settings.
Setting
Recorder type
Start value
End value
Measured variable
Attenuation
I -> UFL/OFL
Behavior at error
>
<OK>
and
, highlight and confirm Save and quit.
Selection/Values
0 to 20 mA or
4 to 20 mA
(sensor
dependent)
Main variable
Adjoining variable
0 ... 40 mA/s
Error
Limitation
Fixed current
value
Explanation
Minimum spacing:
≤ 5 % of the measuring range
(sensor-dependent)
Main variable designates the
actual measured parameter of
the sensor (e.g. pH, oxygen,
etc.).
Adjoining variable designates an
additional measured parameter
(e.g. temperature).
Speed of change of the output
current (mA/s) in the case of
erratic changes of the input
signal.
Current values outside the range
between Start value and End
value are regarded as an error.
The current output reacts as
specified under Behavior at error
(see below).
The current at the output is
limited to the Start value or End
value.
In the case of system errors and
sensor errors, the current output
delivers the fixed current value
that was specified.
Possible values:
0 ... 21 mA.
ba77244d01
DIQ/S 281
08/2020