Xylem WTW DIQ/S 281 Operating Manual page 31

Universal transmitter for one digital sensor
Hide thumbs Also See for WTW DIQ/S 281:
Table of Contents

Advertisement

DIQ/S 281
ba77244d01
X19
X18
L1
N
100...
240V AC
NETZ/MAINS
L
N
figure 3-14 Line power connection.
The complete assignment of the terminal strip is shown in
section 3.12.
9
Connect phases L and N to the terminal strip. Make sure that the cable
assignment agrees with the specification on the terminal label under the
terminal strip.
10
Tighten the cap nut (pos. 3 in figure 3-13).
WARNING
No free wires are allowed to project into the housing. Oth-
erwise, there is a danger that areas safe to contact could
come into contact with dangerous voltages. Always cut
off any wires that are not in use as closely as possible to
the cable gland.
11
Close the enclosure of the DIQ/S 28X.
08/2020
Installation
31

Advertisement

Table of Contents
loading