Fixed Current Value - Xylem WTW DIQ/S 281 Operating Manual

Universal transmitter for one digital sensor
Hide thumbs Also See for WTW DIQ/S 281:
Table of Contents

Advertisement

DIQ/S 281
Function
ba77244d01
Setting
Io
ti
td
Behavior at error

6.7.3 Fixed current value

The Fixed current value function can be used to test the operativeness of the
instruments connected to the outputs: output different current values to the out-
put and, while doing so, check the behavior of the connected instrument.
Setting
Current output
Inom
The settings for other functions in the Current output menu as, for
example, PID controller and Recorder are retained while the Fixed
current value is carried out.
08/2020
Selection/Values
0 ... 20 mA
0 ... 9999 s
0 ... 9999 s
Fixed current value
Unchanged
Selection/Values
Fixed current
value
0 ... 20 mA
Explanation
Current value on the output if
the measured value equals the
Nominal value
Hold-back time:
Integral part of the controller
(0 = not effective)
Reset time:
Differential part of the
controller
(0 = not effective)
In the case of an error, the
current output supplies the
current value defined in the
Current with error field (any in
the range
0 ... 21 mA).
In the case of an error, the
current on the output remains
unchanged.
Explanation
Using Save and quit, the nominal
amperage that was input as Inom
is output.
The nominal amperage that is
output.
Outputs
111