Maintenance; General; Intervals; Redox Electrodes - Xylem MJK pHix Compact Manual

Ph / redox transmitter
Hide thumbs Also See for MJK pHix Compact:
Table of Contents

Advertisement

Maintenance

General

A pH or redox measuring system requires maintenance with regular

intervals.

Cleaning and buffer adjustment (calibration) is a normal and necessary
maintenance that should be carried out regularly.
Intervals
The interval between buffer adjustments and cleaning depends widely on
the working conditions. Therefore, the easiest way to find the most
appropriate cleaning interval is by feeling one's way.
For instance, start out by cleaning and buffer adjusting every week.
If measurements are unchanged or just slightly changed after buffer
adjustment, the electrode were not dirty or just slightly dirty, and the
cleaning interval can be changed to two weeks and after that, one month.
The intervals are equally reduced if the electrodes are very dirty and the
measurements are significantly changed after buffer adjustment.
There are no general guidelines to cleaning and adjustment, but in most
cases, a monthly adjustment is sufficient.
Before beginning a buffer adjustment, the electrode must be thoroughly
washed:
1. first in running water
2. then, if necessary, in a degreasing fluid, like, for example a 10%
hydrochloric acid solution
3. then, before the electrode is put into the buffer solution, and again in
between every step of the buffer adjustment, rinse with water and dry
with a dry cloth.

Redox electrodes

Like pH electrodes, Redox electrodes require cleaning but not buffer
adjustment.
However, MJK does deliver redox buffers for control of electrode and
transmitter.
32
MANUAL
EN 4.10 pHix® Compact 1904

Advertisement

Table of Contents
loading

Table of Contents