General Course When Calibrating, Cleaning, Servicing Or Repairing An Iq Sensor - Xylem WTW DIQ/S 281 Operating Manual

Universal transmitter for one digital sensor
Hide thumbs Also See for WTW DIQ/S 281:
Table of Contents

Advertisement

Operation
60
figure 4-11 List of all components
6
Exit the Status of output channels window with
4.8
General course when calibrating, cleaning, servicing or
repairing an IQ sensor
When an IQ sensor is calibrated, cleaned, serviced or repaired, the mainte-
nance condition for the IQ sensor should always be switched on.
In the maintenance condition
 the system does not react to the current measured value or the condition of
the selected IQ sensor
 linked outputs are frozen
 IQ sensor errors do not prompt changes in the conditions of linked outputs.
The maintenance condition for IQ sensors is automatically activated
 during calibration. After calibration the IQ sensor remains in the maintenance
condition until the maintenance condition is switched off manually (see
section 4.8.3)
 during a compressed air cleaning cycle.
Therefore, keep to the following course when you calibrate, clean, service or
repair an IQ sensor.