Log Book - Xylem WTW DIQ/S 281 Operating Manual

Universal transmitter for one digital sensor
Hide thumbs Also See for WTW DIQ/S 281:
Table of Contents

Advertisement

Operation
Structure of the log
book
54
In the case of errors, immediately open the detailed message text in
the log book and perform the recommended actions. If the actions
have been performed, mark the message as read (see
section 4.5.3).

4.5.2 Log book

The log book is a list with all the messages from all modules. The log book
contains up to 1000 entries. If there are more than 1000 messages, the oldest
entries are deleted.
New messages can be recognized by the flashing information symbol or error
symbol. The log book always contains the latest message in the first position.
They do not yet have a checkmark in the status field.
The flashing of the info or error symbol only stops after all detailed message
texts in the log book have been opened and marked with a checkmark ( ) (see
section 4.5.3).
1 2
3
figure 4-9 Log book of entire system
1 Message category (error or info symbol)
2 Module that triggered the message.
SYS system (Universal Transmitter, controller)
S01 IQ sensor (number 01)
D01 DIQ/S 28X output (number 01)
3 Message code
4 Date and time of the message
5 Status field of the message
Message was acknowledged
No checkmark Message was not acknowledged
4
5
ba77244d01
DIQ/S 281
08/2020

Advertisement

Table of Contents
loading