Maintenance; Intervals; Redox Electrodes - Xylem MJK pHix Compact V3 Manual

Next generation
Table of Contents

Advertisement

Maintenance

A pH or redox measuring system requires maintenance with regular intervals.
Cleaning and buffer adjustment (calibration) is a normal and necessary
maintenance that should be carried out regularly.

Intervals

The interval between buffer adjustments and cleaning depends widely on the
working conditions. Therefore, the easiest way to find the most appropriate
cleaning interval is by checking for drift or deviation and feeling one's way.
For instance, start out by cleaning and buffer adjusting every week.
If measurements are unchanged or just slightly changed after buffer
adjustment, the electrode were not dirty or just slightly dirty, and the cleaning
interval can be changed to two weeks and after that, one month.
The intervals are equally reduced if the electrodes are very dirty and the
measurements are significantly changed after buffer adjustment.
There are no general guidelines to cleaning and adjustment, but in most cases,
a monthly adjustment is sufficient. Before beginning a buffer adjustment, the
electrode must be thoroughly washed:
1.
first in portable water
2.
then, if necessary, in a degreasing fluid, like, for example a 10%
hydrochloric acid solution
3.
then, before the electrode is put into the buffer solution, and again in
between every step of the buffer adjustment, rinse with water and dry
with a dry cloth.

Redox electrodes

Like pH electrodes, Redox electrodes require cleaning but not buffer
adjustment. However, MJK does deliver redox buffers for calibration of
electrode and transmitter.
If the transmitter is deviating from the reference values when put into the buffer
solution, the electrode must be exchanged.
Maintenance /Intervals
pHix® Compact V3. NG (next generation)
EN 4.10 Version 2103
30

Advertisement

Table of Contents
loading

Table of Contents