Fix the impact protection plate (Pos. 4) to the back of the DIQ/S 28X
with the bolts (Pos. 5).
3.7

Electrical connections: General instructions

All electric cables are fed from below through openings prepared in the enclo-
sure of the DIQ/S 28X and the DIQ modules. Cable glands with different clamp-
ing ranges are included with the DIQ/S 28X to provide sealing between the
cable and enclosure as well as for strain relief. Select the suitable cable gland
for the relevant cable type and cable diameter:
 Small, clamping range 4.5 to 10 mm.
This cable gland is suitable for all cables (earth cable after stripping the outer
insulation, see section 3.6) and sensor connection cables.
 Large, clamping range 7 to 13 mm.
This cable gland is required for cable sheaths with an outside diameter of
more than 10 mm and is screwed into the enclosure via an extension piece.
If necessary, you can order more big cable glands (see
chapter 10 A
08/2020
sealing ring 20 x 15 x 1 mm
cable gland M16
blind plug
sealing ring 20 x 15 x 1 mm
extension piece M16/M20
sealing ring 24 x 19 x 2 mm
cable gland M20
CCESSORIES AND OPTIONS
Installation
).
27