DIQ/S 281
ba77244d01
figure 5-6
5.6
Alarm settings
5.6.1 General information
Under this menu item you can specify reactions on certain alarm events.
An alarm event is when a certain measured value (limiting value) of a sensor is
exceeded or undercut.
Alarm events can be relayed as follows:
As a message on the display
As a relay action
An alarm message via display or relay cannot be acknowledged or
switched off. An alarm only disappears if the cause for the alarm
was eliminated or the Alarm settings were changed or erased.
5.6.2 Setting up / editing alarms
1
Open the Einstellungen/Settings menu with
<
2
Using
settings.
The Alarm link overview dialog window opens.
Alarms that have already been set up have entries in the Sensor
column.
08/2020
Edit list of outputs -> enter a name
>
<OK>
and
, select and confirm the menu item, Alarm
Settings/setup
<S>
.
75