Delay Trigger - Rigol DS8000-R Series User Manual

Digital oscilloscope
Hide thumbs Also See for DS8000-R Series:
Table of Contents

Advertisement

RIGOL

Delay Trigger

In Delay trigger, you need to set Source A and Source B. The oscilloscope triggers
when the time difference (△T) between the specified edges of Source A (Edge A) and
Source B (Edge B) meets the preset time limit, as shown in the figure below.
Note:
Edge A and Edge B must be two neighboring edges.
Source B
Trigger Type:
Click Type to select "Delay". Then, the current trigger setting information is
displayed at the upper-right corner of the screen, as shown in the figure below.
Source A:
Click SourceA to select CH1-CH4 as the trigger source of Source A. For details,
please refer to descriptions in "Trigger Source". The current trigger source is
displayed at the upper-right corner of the screen.
Note:
Only when we select the channel (that has been input with signals) as the
trigger source, can we obtain a stable trigger.
Edge A:
Click EdgeA continuously to select the trigger edge type of Source A in Delay trigger.
It can be set to "Rising"
Source B:
Click SourceB to select CH1-CH4 as the trigger source of Source B. For details, refer
to descriptions in "Trigger Source". The current trigger source is displayed at the
upper-right corner of the screen.
Note:
Only when we select the channel (that has been input with signals) as the
trigger source, can we obtain a stable trigger.
Edge B:
Click EdgeB continuously to select the trigger edge type of Source B in Delay trigger.
5-26
Source A
T
Figure 5-10 Delay Trigger
or "Falling"
Chapter 5 To Trigger the Oscilloscope
Edge A=
Edge B=
.
DS8000-R User Guide

Advertisement