Delay Trigger - Rigol MS05000-4CH User Manual

Digital oscilloscope
Table of Contents

Advertisement

Chapter 5 To Trigger the Oscilloscope
Edge B:
Press EdgeB continuously to select the trigger edge type of Source B in Delay trigger.
It can be set to "Rising"
Trigger Condition:
Press When to set the time limit condition of Delay trigger.
>: triggers when the time difference (△T) between the specified edges of
Source A and Source B is greater than the preset time lower limit. Press Lower
to set the delay time lower limit in Delay trigger.
<: triggers when the time difference (△T) between the specified edges of
Source A and Source B is smaller than the preset time upper limit. Press Upper
to set the delay time upper limit in Delay trigger.
<>: triggers when the time difference (△T) between the specified edges of
Source A and Source B is greater than the lower limit of the preset time and
smaller than the upper limit of the preset time. Press Upper to set the delay
time upper limit in Delay trigger. Press Lower to set the delay time lower limit in

Delay trigger.

Note:
><: triggers when the time difference (△T) between the specified edges of
Source A and Source B is smaller than the lower limit of the preset time or
greater than the upper limit of the preset time. Press Upper to set the delay
time upper limit in Delay trigger. Press Lower to set the delay time lower limit in
Delay trigger.
Note:
Trigger Mode:
In the trigger control area (Trigger) on the front panel, press Mode to quickly switch
the current trigger mode. For details, refer to descriptions in "Trigger Mode".
Trigger Parameter Setting:
Set the trigger parameters (trigger holdoff and noise rejection) under this trigger
type. For details, refer to descriptions in "Trigger Holdoff" and "Noise
Rejection".
Trigger LEVEL/Threshold Level:
Rotate the Trigger
to "Trigger LEVEL/Threshold Level". The current trigger level/threshold level
value is displayed at the upper-right corner of the screen.
MSO5000 User's Guide
nd Quality Products Online at:
or "Falling"
The lower time limit must be smaller than the upper time limit.
The lower time limit must be smaller than the upper time limit.
LEVEL knob to adjust the trigger level/threshold level. Refer