Chapter 6 Operations and Measurements
Click More → Label to set the label for math operation results. For details, refer
⚫
to descriptions in "Math Operation Label".
Click More → Color Grade to enable or disable the color grade display of the
⚫
operation results. When enabled, different colors are displayed on the screen to
indicate the times of data acquisition or acquisition probability.
Click More → Reset ColorGrade to clear the color grade display and display
⚫
the color grade again.
Division
Divides the waveform voltage values of signal source A by that of source B point by
point and displays the results. It can be used to analyze the multiple relationships of
waveforms of two channels.
Note:
When the voltage of signal source B is 0 V, the division result is treated as 0.
Click Math → Math1 → Operator to select "A÷ B":
Click Operation continuously to enable or disable the division operation
⚫
function.
Click SourceA and SourceB respectively to select the channel for Source A and
⚫
Source B. The available channels for Source A and Source B are CH1-CH4 and
Ref1-Ref10.
Note:
The source of Math2 can be CH1-CH4, Ref1-Ref10, or Math1; the source
of Math3 can be CH1-CH4, Ref1-Ref10, Math1, or Math2; the source of Math4
can be CH1-CH4, Ref1-Ref10, Math1, Math2, or Math3; only the channels that
have been enabled currently can be selected.
Click Offset to set the vertical offset of the operation results. For setting
⚫
methods, refer to the descriptions in "Parameter Setting Method".
Click Scale to set the vertical scale of the operation results. For setting methods,
⚫
refer to the descriptions in "Parameter Setting Method".
Click More → AutoSetting to adjust the vertical scale and offset of the
⚫
operation results to the optimal value according to the current configuration, so
as to better observe.
Click More → Invert to enable or disable the inverted display function of the
⚫
waveform.
Click More → Expand to expand or compress the measurement waveform
⚫
around the "Center" or "GND".
Click More → Label to set the label for math operation results. For details, refer
⚫
to descriptions in "Math Operation Label".
Click More → Color Grade to enable or disable the color grade display of the
⚫
operation results. When enabled, different colors are displayed on the screen to
indicate the times of data acquisition or acquisition probability.
Click More → Reset ColorGrade to clear the color grade display and display
⚫
DS8000-R User Guide
RIGOL
6-5